GE
D
IRECTION FQ091019, REVISION 2 VIVID Q N SERVICE MANUAL
3-20 Section 3-7 - Completing the Hardware Installation
Section 3-7
Completing the Hardware Installation
The hardware installation procedures described in this section provide instructions and information for
the possible scenarios, as follows:
Vivid q N scanner but no SafeLock Cart - connections direct to the system
Vivid q N scanner with SafeLock Cart
Vivid q N scanner but no SafeLock Cart - connections via USB hub
NOTE: If required, it is possible to disable access to all Vivid q N scanner USB ports (including the
CD/DVD drive) in order to limit system vulnerability. This is implemented from the System
Configuration window, via the Service tab.
Activate the USB External Media disabled checkbox (as shown in XREF Figure 3-22 below), then
press the F2 or Esc key to save the changes.
NOTE: After disabling all Vivid q N scanner USB ports, all peripherals that are used by connecting to
a USB port (and also the CD/DVD drive) will not be accessible until such time as access to USB
External Media is re-enabled - i.e., the USB External Media disabled checkbox (shown above)
is unchecked.
Figure 3-12 Disabling USB External Media