IRECTION FQ091019, REVISION 2 VIVID Q N SERVICE MANUAL
Chapter 3 - System Setup 3-3
3-2-3 Safety Reminders
NOTE: The Vivid q N User Manual should be fully read and understood before operating the unit. Keep
the manual near the unit for reference.
WHEN USING ANY TEST INSTRUMENT THAT IS CAPABLE OF OPENING THE
AC GROUND LINE (I.E., METER’S GROUND SWITCH IS OPEN), DO NOT TOUCH
THE UNIT!
Two people are required to unpack the SafeLock Cart, as it is heavy. Two people are always
required whenever a part weighing 19kg (35 lb.) or more must be lifted.
If the unit is very cold or hot, do NOT turn ON power to the unit until it has had sufficient time to
acclimate to its operating environment.
To prevent electrical shock, connect the unit to a properly grounded power outlet.
Do NOT use a three-prong to two-prong adapter, as this defeats safety grounding.
To ensure proper grounding, connect this equipment to a receptacle marked “HOSPITAL ONLY”
OR “HOSPITAL GRADE”.
Do NOT wear the ESD wrist strap when you work on live circuits where more than
30 V peak is present.
Do NOT operate the unit unless all board covers and frame panels are securely in place, to
ensure optimal system performance and cooling. (When covers are removed, EMI may be
present).
ACOUSTIC OUTPUT HAZARD
ALTHOUGH THE ULTRASOUND ENERGY TRANSMITTED FROM THE VIVID Q N PROBE IS
WITHIN AIUM/NEMA STANDARDS, AVOID UNNECESSARY EXPOSURE. ULTRASOUND
ENERGY CAN PRODUCE HEAT AND MECHANICAL DAMAGE.
When the SafeLock Cart is connected to the wall outlet and the main circuit breaker is in the ON
position, the AC Box Fan is operated, even if the system is turned OFF.