IRECTION FQ091019, REVISION 2 VIVID Q N SERVICE MANUAL
3-132 Section 3-10 - Connectivity Setup
3-10-8 eVue Dataflow Configuration
3-10-8-1 Introduction
It is possible to configure the Vivid q N portable ultrasound scanner to allow the user to send images or
loops to a remote PC, in almost real time. When pressing Store, the loop will save not only to the local
archive (in raw-data format), but will also be transmitted to a designated PC (“e-Vue PC”) on the network
(in compact MPEG format), and will appear at the review PC within a very short time.
This feature proves very useful in the OR scenario, where consultation with another doctor in the
hospital in real-time is required.
3-10-8-2 Preparations
Before configuring the Vivid q N for eVue dataflow, it is necessary to decide upon a specific PC in the
hospital as being the e-Vue PC station. The service technician then configures the
Vivid q N to send images to that specific e-Vue PC, as described below.
3-10-8-3 Configuring the Vivid q N for eVue Dataflow
1.) Verify which PC in the hospital/institute has been designated as the e-Vue station.
2.) Make sure the Vivid q N is connected to the Network via cable or wireless.
3.) On the Vivid q N scanner, press the Config button.
The System Configuration window opens.
4.) Click Connectivity.
The Connectivity dialog box opens.
5.) Select the Dataflow Tab, as shown in Figure 3-140 on page 3-133.