EasyManua.ls Logo

GE Vivid q N - Lower Cover (Rear) Replacement Procedure

GE Vivid q N
698 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE
D
IRECTION FQ091019, REVISION 2 VIVID Q N SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-73
8-4-6 Lower Cover (Rear) Replacement Procedure
8-4-6-1 Tools
Appropriate Phillips screwdriver.
8-4-6-2 Preparations
Shut down the Vivid q N ultrasound unit, as described in Chapter 3 - System Setup.
8-4-6-3 Lower Cover (Rear) Removal Procedure
Remove six retaining screws of the lower cover (rear) and while supporting the cover with one hand,
gently remove the cover - see Figure 8-86).
8-4-6-4 Lower Cover (Rear) Installation Procedure
1) Return the cover to the correct position on the rear of the SafeLock Cart - see Figure 8-86.
2.) While holding the cover to prevent it from falling, fasten the retaining screws of the lower cover
(rear).
3) Proceed to perform the following functionality tests:
- "SafeLock Cart - Grounding Continuity" on page 10-22
- "SafeLock Cart - Chassis Current Leakage Test" on page 10-24
Figure 8-86 Lower Cover (Rear) Removal
Retaining Screw

Table of Contents