IRECTION FQ091019, REVISION 2 VIVID Q N SERVICE MANUAL
8-66 Section 8-4 - SafeLock Cart Components Replacement
8-4-2 Upper Cover (Rear) Replacement Procedure
8-4-2-1 Tools
Appropriate Phillips screwdriver.
8-4-2-2 Preparations
Shut down the Vivid q N ultrasound unit, as described in Chapter 3 - System Setup.
8-4-2-3 Upper Cover (Rear) Removal Procedure
Remove the retaining screws of the upper rear cover and while supporting the cover with one hand,
carefully remove the cover (see Figure 8-77).
8-4-2-4 Upper Cover (Rear) Installation Procedure
1.) Return the cover to the correct position on the rear of the SafeLock Cart (see Figure 8-77).
2) While holding the cover to prevent it from falling, return and fasten the two upper screws
(see Figure 8-77).
3.) Fasten the four remaining screws that secure the upper cover (rear).
4) Proceed to perform the following functionality tests:
- "SafeLock Cart - Grounding Continuity" on page 10-22
- "SafeLock Cart - Chassis Current Leakage Test" on page 10-24
Figure 8-77 Upper Cover (Rear) Removal