IRECTION FQ091019, REVISION 2 VIVID Q N SERVICE MANUAL
Chapter 3 - System Setup 3-113
9.) In the Select Tag drop down menu select "00080060 Modality".
10.)Type US in the Value text box and click on "Add to List".
The Search Criterias window is updated (see Figure 3-118).
11.)Click OK to close the window.
12.)From the Worklist Properties Dialog, select the IP-Address down-arrow to choose the Worklist
Server from the pull-down menu.
NOTE: It is not possible to change the setting in the IP-Address field by editing it. See description
below, starting with step 13, to change the setting.
Figure 3-119 Search Criterias Window - Updated