GE
D
IRECTION FQ091019, REVISION 2 VIVID Q N SERVICE MANUAL
3-124 Section 3-10 - Connectivity Setup
3-10-6-3 Query/Retrieve Verification
Follow the Steps Below to do a First Test (Ping) of the Connection
1.) Select the Smiley button to Ping the server.
2.) If the network connection to the server is OK, it will be illustrated by a smiling Smiley .
A sad Smiley indicates that the network connection is failing.
Typical cause: Network cable not connected.
3.) When ready, click OK to close the DICOM Query/Retrieve properties and save changes.
Figure 3-132 DICOM Query/Retrieve Properties
SMILEY
OK
VERIFICATION
CHECK BOX

Table of Contents