GE
D
IRECTION FQ091019, REVISION 2 VIVID Q N SERVICE MANUAL
Chapter 6 - Service Adjustments 6-3
The Test Pattern in the Calibration Screen (refer to Figure 6-1 on page 6-1) may be used for Advanced
Settings. (Press Alt + Q to access the Test Pattern).
6-1-2-4 Blue Tint Adjustment Procedure
This will make the white shades shift a bit towards bluish.
• To adjust the Blue Tint: hold down <ALT>, then use the < and > keys, as required.
Repeatedly pressing < will decrease the blue tint; repeatedly pressing > will increase the blue tint.
To assist in making blue tint adjustments, an adjustment bar is displayed below the screen:
6-1-2-5 Blue Tint “2” Adjustment Procedure
NOTE: Systems with BT11 installed software have a Blue Tint “2” option that enables blue tint
enhancement. This option can be switched off if preferred by the user.
NOTE: This option applies to the following LCD monitor installations:
System: Vivid i n
Monitor: BOE HYDIS-1
NOTE: On older Vivid i/q systems where the IDTECH-1 monitor is supplied, the Blue Tint “2” option is
not applicable.
NOTE: The Blue Tint “2” option is incompatible with IDTECH-1/2 monitors, therefore, on these
systems, the Blue Tint “2” option must be turned off.
To identify which monitor is installed on the system:
1.) Log on as Administrator.
2.) Press Config > About > and select the HwVersion tab.
The monitor type appears under Module Name and the part number indicates the monitor type
- see Figure 6-6.
Figure 6-5 LCD Display - Blue Tint Adjustment