Chapter 8 - Replacement Procedures 8-5
8-2-2-5 Battery Installation Procedure
1) With the Vivid q N lying face down (refer to Figure 8-5 on page 8-4), hold the unit steady with one
hand while sliding the battery sideways into the battery compartment.
2.) Make sure the battery is properly inserted all the way, before pushing the locking latch to the closed
position.
3.) Turn the Vivid q N over, release the front latches and open the LCD cover.
4.) Reconnect the AC adapter and all cables to the Vivid q N portable ultrasound scanner.
5.) Proceed to perform the following functionality tests:
- "SafeLock Cart - Grounding Continuity" on page 10-22 - only if the Vivid q N system is
currently mounted on the SafeLock cart
- "SafeLock Cart - Chassis Current Leakage Test" on page 10-24
- "Calibration Tests" on page 7-25
- "Battery Parameters Test" on page 7-102
WARNINGWARNING
THE VIVID Q N BATTERY HAS A SAFETY DEVICE.
DO NOT ATTEMPT TO DIS-ASSEMBLE OR ALTER THE BATTERY!
ALWAYS OBSERVE THE PRECAUTIONS LISTED IN "Vivid q N Battery Safety" on page
1-15.
CAUTION
Use only batteries approved by GE as suitable for use with the Vivid q N portable ultrasound
scanner