5.9.5 pH Sensor
Connect the pH sensor
Set the calibration tolerance
Determine the Calibration value 1 and then the Calibration value 2 and confirm with
RETURN
To calibrate the Calibration value 1: Immerse the sensor in the calibration liquid
suitable for your specified calibration value and press RETURN
Leave the sensor in the calibration liquid until the signal tone
The sensor is now calibrated to Calibration value 1
Dry the sensor and continue with the calibration of the Calibration value 2
If no errors have occurred, save the calibrated values by selecting Yes and confirm
with RETURN
Check the calibration as described below
The pH sensor is now ready for use. Based on your calibration, ProfiLux 4 can now determine
the pH value.
5.9.6 Redox-Sensor
For the redox sensor calibration, you need a so-called Null-plug, which is supplied with
ProfiLux 4 as well as our redox expansion cards (PLM Redox). The plug must be plugged onto
the sensor connection of the ProfiLux 4 (corresponding white BNC socket on the back of the
housing).
Set the calibration tolerance
Determine the Calibration value 2 and confirm with RETURN
Plug the Null- plug ->RETURN
Wait for the signal sound
Disconnect the Null- plug and then connect the Redox sensor ->RETURN
Immerse the sensor in the Calibration liquid to suit your specified Calibration value -
>RETURN
Wait for the signal sound
If no errors have occurred, save the calibrated values by selecting Yes and confirm
with RETURN
Check the calibration as described below
The Redox sensor is now ready for use. Based on your calibration, ProfiLux 4 can determine
the Redox value.

Table of Contents

Other manuals for GHL ProfiLux 4

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GHL ProfiLux 4 and is the answer not in the manual?

GHL ProfiLux 4 Specifications

General IconGeneral
ModelProfiLux 4
CategoryController
Power Supply12V DC
pH Ports2
Redox Ports1
Conductivity Ports1
Level Ports4
Digital Inputs4
12V Outputs4
Real-time clockYes, with battery backup
Digital Outputs4
ExpandableYes
ConnectivityUSB, Wi-Fi, LAN
DisplayGraphic LCD display
Expansion Slots4
Processor32-bit
Channels16
Operating Temperature0°C to 50°C