WGM - 28 November 2007 - Edition 01
PAT700 Total Organic Carbon Analyzer – Installation Page 31
Anatel Operator Manual
Example:
Read command: “HR”
Read format:
HH:mm:SS HH:mm:SS u u u u s u u u u u u u u u u u u
HH:mm:SS u u f u u u u u u f
Response:
00:01:00 00:00:00 50 0 1 1 Sensor 0 1 0 3 0 1 0 0 0 0 1 1
00:01:00 25 00 1.000 000 020 005 095 0 0 1.000
Write format:
HR HH:mm:SS HH:mm:SS u u u u s u u u u u u u u u u u u
HH:mm:SS u u f u u u u u u f
Write command:
HR 00:01:00 00:00:00 50 0
1 1 Sensor 0 1 0 3 0 1 0 0 0 0 1 1
00:01:00 25 00 1.000 000 020 005 095 0 0 1.000
Example:
Read command: “HH”
Response:
UVLamp1: 4 4300 1.000000 Good
UVLamp2: 0 4300 1.000000 Good
Example:
Read command: “HC”
Response: “1.000”
Write format: “HC f”
Write command: “HC 0.95”
Example:
Read command: “HT”
Response: “1.000”
Write format: “HT f”
Write command: “HT 1.05”
“HH”
Get number of times UV lamp is turned on, hours left, voltage and status of both UV lamps
“HC”
Set/get user-calibrated
conductivity slope
Slope range should be between 0.9 and 1.1.
“HT”
Set/get user TOC slope Slope range should be between 0.85 and 1.15.