Page 60 Setup – PAT700 Total Organic Carbon Analyzer
Operator Manual Anatel
WGM - 28 November 2007 - Edition 01
General tab
Figure 5-12 Security settings dialog box, general tab
Enable security checkbox allows you to enable or disable the analyzer’s security system.
When security is enabled, you are required to sign on prior to making changes in the analyzer,
and only the administrator may access the security settings. When security is disabled, you
are not required to sign on. Disabling the analyzer’s security system disables the audit trail
and all controls on this dialog box except the enable data logging checkbox. The default
setting is disabled.
Enable data logging checkbox allows you to enable or disable data logging. When enabled,
all data (up to 5000 records) are automatically stored in the analyzer’s internal data buffer.
When disabled, no data are saved. The default setting is enabled.
Printer icon prints the security settings.
Export icon opens the export settings dialog box. See Figure 5-13.
Return icon returns you to the setup dialog box and saves the changes. Any changes are
noted in the audit trail.
Figure 5-13 Export settings dialog box