Page 82 Bottle Mode – PAT700 Total Organic Carbon Analyzer
Operator Manual Anatel
WGM - 28 November 2007 - Edition 01
C of A value text box allows you to enter the certified analysis value for the
standard that is being used.
Volume text box allows you to enter the volume of water, in mL, for the specified
standards bottle.
Expiration date text box allows you to enter an appropriate expiration date, in ISO
yyyy-mm-dd format, and time, in hh:mm:ss format, for the specified TOC calibration
standard.
Forward icon takes you to the review test setup dialog box. See Figure 6-6.
Cancel icon returns you to the load bottles dialog box.
Figure 6-8 Schedule test dialog box
Date text box allows you to enter the date the test will run. The default value is
tomorrow’s date. You may enter any valid date, in ISO “yyyy-mm-dd” format, that is
within three days from the loading of the standards.
Time text box allows you to enter the time the test will run. The default time is the
current time. You may enter any valid time in hh:mm:ss format.
Back icon takes you to the previous dialog box. All changes made in this dialog box
are retained.
Finish icon takes you back to the home screen. All changes made in any of the
dialog boxes are retained and the test is scheduled to run. Prior to exiting this dialog
box, the analyzer verifies that the scheduled date and time is within three days of
the current date and time. It also verifies that none of the bottles will expire prior to
the scheduled date.
10) When the test running dialog box appears, the analyzer begins running the test.
See Figure 6-9.