EasyManua.ls Logo

Hantek IDSO1070 series - Self Calibration

Hantek IDSO1070 series
77 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Quick Start
User’s Manual 7
1.7 Self Calibration
The self calibration routine lets you optimize the oscilloscope signal path for maximum
measurement accuracy. You can run the routine at any time but you should always run the
routine if the ambient temperature changes by 5 or more. For accurate calibration, power on
the oscilloscope and wait twenty minutes to ensure it is warmed up. To compensate the signal
path, disconnect any probes or cables from the input connectors. Then, access the “Utility ->
Calibrate/Self Calibration option and follow the directions on the screen. The self calibration
routine takes about several minutes.

Table of Contents

Related product manuals