Base: Voltage of the statistical minimum level, Measured over the entire waveform
Middle: Voltage of the 50% level from base to top
RMS: The Root Mean Square voltage over the entire waveform
Amplitude: Amp = Base - Top, Measured over the entire waveform
Mean: The arithmetic mean over the entire waveform
Cycle Mean: The arithmetic mean over the first cycle in the waveform
Positive OverShoot: Positive Overshoot = (Max - Top)/Amp x 100 %, Measured over the entire
waveform
Negative OverShoot: Negative Overshoot = (Base - Min)/Amp x 100 %, Measured over the entire
waveform.
3. Horizontal
Period: Time to take for the first signal cycle to complete in the waveform
Frequency: Reciprocal of the period of the first cycle to complete in the waveform
Rise Time: Time taken from lower threshold to upper threshold
Fall Time: Time taken from upper threshold to lower threshold
+Duty Cycle: Positive Duty Cycle = (Positive Pulse Width)/Period x 100%, Measured of the first
cycle in waveform.
-Duty Cycle: Negative Duty Cycle = (Negative Pulse Width)/Period x 100%, Measured of the first
waveform.
+Pulse Width: Measured of the first positive pulse in the waveform. The time between the 50%
amplitude points
-Pulse Width: Measured of the first negative pulse in the waveform. The time between the 50%
amplitude points
4. Clear Item
Clear all measure items on display screen.
5. Edit Item
Click Measure->Edit Item. User can select Maximum 8 measure options to measure.
Note: The results of the automatic measurements will be displayed on the bottom of the screen.
Maximum 8 results could be displayed at the same time. When there is no room, the next new
measurement result will make the previous results moving left, out of screen.
6. DVM
Click Measure->DVM, the DVM dialog will be showed.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hantek IDSO1070 series and is the answer not in the manual?

Hantek IDSO1070 series Specifications

General IconGeneral
BrandHantek
ModelIDSO1070 series
CategoryTest Equipment
LanguageEnglish

Related product manuals