Theory of Operation
11
Caution: Only Selectatec™ compatible vaporizers with interlock system may be used
with the Modular
+
unit.
Caution: After each exchange of a vaporizer, carry out a system leak test.
Caution: Use cleaning agent sparingly. Excess fluid could enter the machine causing
damage.
Caution: The patient dome of the bellows system cannot be autoclaved. It is not in
contact with the ventilation gas. If soiled, the patient dome should be cleaned
with water and liquid cleaning agent. The unit can be disinfected with a standard
surface-disinfecting agent. Do not use alcohol.
Caution: Do not clean the machine while it is on and/or plugged in.
Caution: Pressing Quit at any time during the calibration procedure will cancel the
session’s settings and reload the previously stored calibration coefficients.
1.6.3 Notes
Note: Unauthorized servicing may void the remainder of the warranty. Check with the
factory or with a local authorized HEYER dealer to determine the warranty status
of a particular instrument.