Preventive Maintenance
85
4.6.2.4 Temperature Sensor Calibration
(Skip to Step 5 if already in the Main menu.)
1 Enter the Service Software by pressing the following keys prior to the end of the startup routine:
a
SET
b
OPT.
c
d
OK
2 Select KEYCODE to enter the diagnostic screens.
3 Enter a valid key code to invoke the service software.
Advance from this window to the main menu by pressing ENTER.
4 Select TEMP.SENS to enter the Temperature sensor screen.
Fig. 37 Temperature Sensor screen