EasyManua.ls Logo

Heyer Modular+ - Internal Flow Sensor Calibration

Heyer Modular+
114 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Preventive Maintenance
92
4.6.2.6 Internal Flow Sensor Calibration
Caution: The three procedures “Proportional Valve Calibration”, “Internal Flow Sensor
Calibration” and “Characteristic of the Proportional Valve” must always be
performed together in the order as in the manual, never individually.
1 Select FLOW.INT to enter the Internal flow sensor screen sensor screen.
Fig. 42 Internal Flow Sensor Calibration Screen
2 Adjust minimum range calibration.
a Select MIN to apply a zero flow.
b Select MIN again to record the value.
3 Adjust maximum range calibration.
a Select MAX to apply the maximum flow.
b Select MAX again to record the value.
4 Select MIN again to close the valve.
5 Select SAVE to store the values into the memory.
6 Select QUIT to return to the main menu.

Table of Contents

Related product manuals