EasyManua.ls Logo

Heyer Modular+ - Fig. 40 Pressure Regulator for the Driving Gas with Test Port

Heyer Modular+
114 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Preventive Maintenance
89
4 Adjust the maximum range calibration.
a Select MAX to open the Proportional valve.
b Turn the P.V.GAIN potentiometer (P2, AVM 3-2) until screen pointer shows 97.
c Select MAX twice to record the value.
5 Measure the gas flow at the drive gas outlet of the docking station.
a Verify that the flow is at least 75 (–5) l/min
b Select MIN to close the proportional valve.
c If the gas flow was within specification, skip the steps 14 and 15 and perform step 16. If the
flow rate was out of specification, go to step 14.
6 High-pressure regulator adjustment.
a Remove the Driving gas pressure source (AIR or O
2
).
b Disconnect the red plug of the test port on top of the pressure regulator for the driving gas.
Fig. 40 Pressure Regulator for the Driving Gas with Test Port
c Attach a pressure gauge to the test port.
d Reconnect the Driving gas pressure source (AIR or O
2
).
e Adjust the high pressure regulator by loosing the locking nut on the regulator and turning the
adjustment screw until 200 ± 0.05 kPa (2 ± 0.05 bar; 29 ± 1 PSI) is displayed on the pressure
gauge. Retighten the locking nut.
f Remove the Driving gas pressure source (AIR or O
2
).
Test Port

Table of Contents

Related product manuals