Preventive Maintenance
95
Tolerance rage for the flow values:
Flow Setting Tolerance
1 l/min: +0 / - 0.1
2 l/min: +0 / - 0.1
4 l/min: +0 / - 0.1
6 l/min: +0 / - 0.1
8 l/min: +0 / - 0.1
10 l/min: +0 / - 0.2
12 l/min: +0 / - 0.2
15 l/min: +0 / - 0.2
20 l/min: +0 / - 0.4
25 l/min: +0 / - 0.4
30 l/min: +0 / - 0.4
35 l/min: +0 / - 0.4
40 l/min: +0 / - 0.4
45 l/min: +0 / - 1
50 l/min: +0 / - 1
55 l/min: +0 / - 1
60 l/min: +0 / - 1
65l/min: +0 / - 1
70 l/min: +0 / - 1
75 l/min: +0 / - 5
6 Select ENTER again.
7 Select SAVE twice.
8 Select QUIT to stop the flow of gas and to return to the main menu.
9 Reinstall the patient module.