82 / 106 HEYER Pasithec, Operator’s manual Rev. 0.2 Draft - 12/09
Message Priority
Cause of Alarm Machine
Mode
When
Checked
(Startup
/ Run)
Operator Action Repair
O
2
Cal Due Low 72 hours has elapsed
since last 21% O
2
Calibration
Runtime
only
Software
Version Error
- - BDU, UI, and
Keyboard software
versions are
inconsistent with
released set.
Startup
only
Call Service
Representative
Service Rep
needed.
Keyboard
Communication
Failure
Low No communication
between UI and
Keyboard.
Startup
and
Runtime
Pressure
Sensor Failure
- - Pressure sensor data
incorrect.
Startup
only
Call Service
Representative
Service Rep
needed.
Inspiration
Sensor Failure
- - Inspiration sensor
data incorrect
Startup
only
Ventilator failure;
switch to bag
mode and
manually
ventilate patient.
Monitoring is not
available.