Rev. 0.2 Draft – 12/09 HEYER Pasithec, Operator’s manual 85 / 106
Message Priority
Condition Disabled
In
Manual
Mode
Operator
Action
Repair
High PEEP Mid Exp pressure 5 cmH
2
O above
PEEP for 2 breaths, or
Exp. Pressure 5 cmH
2
O
above PEEP in pressure
support mode for more than
30 seconds.
N.A.
Negative
Pressure
High Negative Airway Pressure of -
2 cmH
2
O or less for greater
than 4 seconds.
No