6.2 Saving Data
125
3
Chapter 6 Saving & Loading Data
6
When auto save is enabled before starting measurement, data can be automatically
saved to removable storage during or after measurement.
The following types of measurement data can be auto-saved.
Automatic Saving (Waveform Data and Numerical Calculation
Results)
Saved Data Settings
File Ex-
tension
Description
Waveform
Data Only
Waveform
(realtime)
.MEM
During measurement, waveform data is saved in binary format.
Conversion to text (CSV) format can be done later, so we sug-
gest selecting Waveform(realtime) for typical operation.
CSV
(realtime)
2
.CSV3
During measurement, waveform data is saved in text format.
Saved data cannot be reloaded into the HiLogger or by the Log-
ger Utility. However, this format is ideal for loading into spread-
sheets like Excel.
Calculation
Results Only
1
Calc
(post
meas.)
.CSV3
Calculation results are saved after measurement.
When [Repeat]
recording is enabled (On), each measurement's
calculation results are appended to the file.
When [Split Calculation] is enabled (On), calculation results
are appended at each save interval.
Both Wave-
form Data and
Calculation
Results1
Waveform
+Calc
.MEM,
.CSV3
Waveform data is saved in binary format during measurement,
and calculation results are saved at the end of measurement.
CSV +
Calc2
.CSV3,
.CSV3
Waveform data is saved in text format during measurement, and
calculation results are saved at the end of measurement.
Saved data cannot be reloaded into the HiLogger.
1. Calculation must be enabled before starting measurement (p. 143).
*2. The recording interval setting is limited for [C
SV(realtime)] and [CSV+Calc] according to the
number of channels used.
For up to 15 channels (using U1-1 to U1-15), the rec
ording interval must be longer than 50 ms
For up to 30 channels (using U2-1 to U2-15), the
recording interval must be longer than 100 ms
For up to 45 channels (using U3-1 to U3-15), the reco
rding interval must be longer than 200 ms.
For up to 60 channels (using U4-1 to U4-15), the reco
rding interval must be longer than 200 ms.
To perform waveform calculations, the recording interva
l must be longer than 500 ms.
(Pulse, logic, and alarm channels do not impose such
limits.) Also, when opening a CSV file
in a spreadsheet program, the number of rows that can be loaded is limited.
*3. The TXT file extension is app
lied except when [Separator:Comma] is selected on the [Sys-
tem] s
creen.
Select from the listed op-
tions.
Select the [Setting] Screen.
Move to a setting item.
Apply
Open the setting options
for the item to be set.
Setting Procedure
Displays an estimate of the re-
maining time available on the at-
tached CF card/ USB flash drive.