8.1 Operation Settings
152
If a power outage or other power loss causes an interruption in recording (while the LED
on the left side of START is lit), you can automatically resume recording when the power
is restored. If you are using triggers, the triggers are restored to the [Waiting for trigger]
state.
Start Backup
(
Auto-
Resume)
Key Setting Procedure
Select the [System]
Screen.
Move to a setting item.
Apply
Open the setting options
for the item to be set.
Select from the listed op-
tions.
1
2
8.1 Operation Settings
Using the Auto-Resume Function (Resume After Power Restora-
tion)
When Auto-Resume is enabled, measurement data that was in internal
memory before the outage is lost when measurement resumes. To retain
measurement data from before an outage, disable Auto-Resume (set to
Off).
To retain measurement data from before an outage and resume measuring
automatically afterwards, enable Auto Save (p. 125).
Setting options:( : default setting)
Off
Do not use the Auto-Resume Function.
On Use the Auto-Resume Function.