1 Choose the input unit and channel(s) to serve as
alarm criteria, and select the checkbox [ ].
2 (If alarm criteria settings are disabled)
Select [Alarm on.].
ALM1 to ALM4 outputs are all enabled [ ].
1 Select [PLS&ALM].
2 Choose an alarm channel (ALM1 to ALM4) for
output, and select the checkbox [ ].
To enable ALM1 to ALM4 outputs, on the setting screen for
channels 1-1 to 4-5 and P1 to P8, move the cursor to
[Alarm on.], select the checkbox [ ], and
press ENTER.
ALM1 to ALM4 are enabled and disabled together. ALM1
cann
ot be enabled by itself.
3 Select whether to reset or hold (latch) alarm out-
puts while measuring.
This setting is also available on the [Gauge+Wave]
Waveform/Numerical screen (p. 110).
4 Enable or disable the alarm beeper by the
[Buzzer] setting.
This setting is also available on the [Gauge+Wave]
Waveform/Numerical screen (p. 110).
Setting options: ( : default setting)
Off
Disable the alarm output signal when alarm
criteria are no longer met.
On Hold the alarm output active state until mea-
surement is finished.
Setting options: ( : default setting)
Off
The HiLogger will not sound an audible beep
during alarm signal output.
On The HiLogger will sound an audible beep dur-
ing alarm signal output.
1
: On
: Off
4
1. Configuring Alarm Output
Select the [CH] Screen.
• Alarm hold and beeper
can also be set on the
Waveform/Numerical
screen ([Gauge+Wave]
display).
• To stop the alarm beeper
while measuring, set the
beeper to [Off] in the
ALM channel settings on
the Waveform/Numeri-
cal screen.
(continues on next page)
2
2. Setting Alarm Criteria for Each Channel
1
: Displayed when the box
is not checked in step 2
of procedure 1.
2
3