1.4 Basic Operation
25
3
1
Chapter 1 Overview
Keys can be disabled to avoid inadvertent operations.
Measurement data, settings, screen images and numerical calculation results can be
save
d.
For details of the saving methods, see "Chapter 6 Saving & Loading Data" (p. 121).
(Default settings)
Saving Method: [Select & Save] (Press SAVE to select and save data.)
Disabling Key Operations (Key-Lock Function)
Press and hold the right and left cursor keys for
three seconds to lock the other keys, and repeat
to unlock.
Performing Zero Adjustment
Saving Data
With the Waveform/Numerical or [CH] screen displayed,
press the up and down cursor keys simultaneously.
See: "2.7 Compensating for Input Circuit Offset (Zero Adjustment)"
(p. 48)
Insert a CF card (or
USB flash drive).
Press SAVE.
(A dialog appears.)
Select the save destination
and item to save, and press
ENTER.
Select [Yes] in the confirma-
tion dialog, and press ENTER
again.
(Data is saved on the selected re-
movable storage.)