Appendix 2 Error Messages and Remedial Actions
A2
Error messages consist of either "Error" or "Warning" displays.
A screen message appears wheneve
r an error occurs. In either case, take the reme-
dial action indicated.
To clear an error message, press ENTER or ESC.
Appendix 2 Error Messages and Remedial
Actions
Error Messages
Message Remedial Action
001
Failed to back up the waveform.
Data in internal memory will be lost in about 30
mi
n. after power outage. For long-term measure-
ments, Auto Save to CF card or etc. is recom-
mended to prepare for power outages.
See: "What happens to data in a power outage?"
(p .123)
004
Battery low. Cannot access to the
waveform backup and Media.
Low battery charge: recharge or replace with a
fresh battery pack.
005
FPGA error.
Incompatible FPGA firmware version. Try updat-
ing
the firmware, and if the error persists,
submit the HiLogger for repair.
008
FPGA PLL error. Turn off and on the
power.
Measurement stopped due to FPGA internal PLL
fau
lt. Turn power off and on, check ambient con-
ditions (temperature and noise), and try measur-
ing again.
009
Turn off and connect UNIT in numer-
ical order.
If input unit 4 is used while un
it 3 is not, move con-
nections to unit 3 so there is no gap between in-
put units.
010
UNIT recognize error.
An input UNIT could not be properly recognized.
Contact your dealer or Hioki representative, as it
may be damaged.
011
The time of the clock was initialized.
Reset the clock. If the problem remains, have the
HiLogger repaired.
012
Internal Memory recognize error.
Internal memory could not be properly recog-
nized. Contact your dealer or Hioki representa-
tive, as it may be damaged.