Setting Example
Unit: PLS&ALM, Channel: P1,
Input: Count, Count Mo: Add
Filter: On
Scaling
Cond: Dec, Unit: kW, 1kW=50,000 Pulse
Disp Span
Type: Up/Lwr, Upper: 10000, Lower: 0
Make input channel settings on the [CH] screen.
Measurement Hints
• Enable scaling to read measured values directly as electrical energy [kWh]. The display range
changes automatically when scaling is enabled, so set the upper and lower display limits after
enabling scaling.
• When the watt-hour meter uses mechanical (relay) contacts for output, chattering effects can
be removed by enabling the filter.
0 kW
10000 kW
When set as shown at the left,
the vertical axes appear like this.
1
2
Select
Apply
• Data is recorded for the specified recording time and saved
to CF card. Recording stops 30 days after starting.
• To stop recording early, press STOP.
Press START.