10.2 Timing Chart
186
The measurement current will not be stopped for measurement ranges of 10 k and
greater (continuous application).
Do not apply a TRIG signal while measuring (when the INDEX signal is OFF) (the signal
will be retained only once). The TRIG signal is held during self-calibration.
See: "Self-calibration timing" (p. 191)
When changing settings such as measurement range, allow processing time (100 ms)
before applying a TRIG signal.
When not displaying the Measurement screen and while error messages are being dis-
played, input signals are disabled. However, the PRINT signal is valid on the Statistical
Calculation Results screen.
The judgment result and BCD output are finalized before the EOM signal changes to ON.
However, if the controller’s input circuit response is slow, it may be necessary to insert
wait processing after the EOM signal’s changing to ON is detected until the judgment
results are acquired.
TRIG
ON
OFF
t1
Stopped
Delay
t4
INDEX
EOM
OFF
OFF
OFF
ON
ON
ON/OFF
t0
t3
OFF
OFF
OFF
Judgment result/ BCD
Positive-direction
application
Judgment result/ BCD: HI, IN, LO, ERR, BCDm-n, RNG_OUT0 to 3
When OVC is ON
t3
t2
Negative-direction
application
Delay
t2
Measurement
processing
Measurement
currents
Measure-
ment
Contact check
Measure-
ment
Positive-
direction
application
Check in
progress
Check in
progress