Chapter 13 Specifications
252
∗1 Negative values can be up to -10% full scale.
The maximum display range is 9,999,999 dgt. or 9 G.
(If the maximum measurement range is exceeded, the over-range display will be shown even if the value is less
than or equal to the maximum display range.)
∗2 • When LP: OFF, 0.001% f.s. = 10 dgt. However, when using the 100 M range or greater with the 100 M range
high-precision setting off, 0.01% f.s. = 1 dgt. With LP: ON, 0.001% f.s. = 1 dgt.
• Measurement accuracy is the accuracy after zero-adjustment. When not performing zero-adjustment, the value
indicated under [Additional accuracy without 0ADJ] is added.
• Values in parentheses on the second row apply when OVC is on. Only OVC on values are provided in the “Low
Power: ON” table.
• During temperature correction, the following value is added to the resistance measurement accuracy rdg. error:
t
0
: Standard temperature [°C]
t : Current ambient temperature [°C]
Δt : Temperature measurement accuracy
α
t0
: Temperature coefficient [1/°C] at t
0
∗3 Measurement current accuracy is ±5%
When using the 1,000 range or lower with an external trigger source or with continuous measurement off (non-
free-run), the measurement current is only applied from the start of measurement (TRIG = ON) to the end of mea-
surement (INDEX = ON). The measurement current is stopped at all other times. When using the internal trigger
source with continuous measurement on (free-run), the measurement current is stopped while the contact check
indicates an error.
∗4 When using an external trigger source or when continuous measurement is off (non-free-run), the open voltage is
limited to 20 mV or less from 1 ms after the completion of measurement (INDEX = ON) until the start of the next
measurement (TRIG = ON).
∗5
When the contact check function is off (when the contact check function is on, 300 mV)
Measurement time (unit: ms) tolerance: ±10%±0.2 ms
-
α
t0
Δt
1+
α
t0
× (t + Δt - t
0
)
× 100 [%]
When using the internal trigger source with continuous measurement on (free-run): Time of 1 measurement
in the measurement target connected state
When using an external trigger source or with continuous measurement off (non-free-run): From trigger input
until EOM turns on
A: Trigger detection time (unit: ms)
B: Contact improvement time (unit: ms)
C: Delay setting (unit: ms)
OVC
*1
Measurement time
OFF
ON
OVC
*1
Measurement time
OFF
ON
TRIG logic setting Time
ON edge
0.1
OFF edge
0.3
Contact improver function Time
OFF
0.0
ON
0.2
Time
Varies with setting.