14.1 Troubleshooting
293
14
6-4 Measured values
are unstable.
See No. 2-1 above.
p.285
6-5 The measured
value differs from
the expected
resistance value.
The wrong channel is being mea-
sured.
Check the current channel and
the channel setting.
p.152
The wiring is shorted.
Check the wiring.
The wiring resistance is high.
For 2-wire connection, the wir-
ing resistance affects the mea-
sured value as-is. Perform zero-
adjustment.
p.163
Measurement leads are connected
to the measurement terminals on
the front of the instrument.
Do not connect measurement
leads to the measurement termi-
nals on the front of the instru-
ment when using the
multiplexer.
p.142
6-6 No measured
value is dis-
played.
The wrong channel is being mea-
sured.
Check the current channel and
the channel setting.
p.152
Display NO UNIT The set unit number and the unit
number in which the Z3003 is
installed differ.
Check the settings and rear of
the unit.
p.148
p.42
The connected device is set to an
external device.
Set the connected device to the
RM3545.
p.155
The relays are worn.
Perform the Multiplexer Unit self-
test. If it yields a FAIL result,
have the Z3003 repaired.
p.166
p.301
The wiring is shorted.
→
Check the wiring.
See No. 2-3 above.
p.287
Fuse blown
→
Check the wiring. If you are still
unable to perform measure-
ment, the internal protective
fuse may have blown. Have
the Z3003 repaired.
p.146
No Issue Items to check Possible causesSolutions See