14.1 Troubleshooting
292
5. Printer issues
6. Multiplexer issues
No Issue Items to check Possible causesSolutions See
5-1 No data is being
printed.
The printer is not connected.
Check whether the connector
has been connected.
Check whether the interface set-
ting is correct.
If using the PRINT signal, see No.
3-3 above.
p.237
p.289
5-2 Printed text is
garbled
The printer and instrument set-
tings do not match.
Check the printer settings again.
No Issue Items to check Possible causesSolutions See
6-1 It is not possible
to switch to the
multiplexer
inputs.
Display ERR:60 Measurement leads are connected
to the measurement terminals on
the front of the instrument.
Do not connect any measure-
ment leads to the measurement
terminals on the front of the
instrument.
If ERR:60 is displayed even
though no measurement leads
are connected, turn off the
instrument and remove the
Z3003. If ERR:60 is no longer
displayed after removing the
Z3003, the Z3003 may be bro-
ken. Have it repaired.
p.148
6-2 Channels cannot
be switched by
operating the
instrument’s
keys.
CH is not being
displayed.
The front terminals are being used
as the measurement terminals.
Set the measurement terminals
to MUX.
p.148
Scan display (set
to list display)
The scan function is set to auto or
step.
Set the scan function to OFF in
order to switch channels with
key operation.
p.148
The set unit number and the unit
number in which the Z3003 is
installed differ.
Check the settings and the unit
on the back of the instrument.
p.148
p.42
RMT is being dis-
played.
The instrument is in remote mode,
in which it is controlled by commu-
nications functionality.
Operate the instrument after
canceling remote mode.
p.230
6-3 Channels cannot
be switched with
EXT I/O.
The MUX signal is not on.
Turn on the MUX signal.
p.180