Appendix 19 Calibration
A39
Appendix
D/A output
Calibration points
*1 Without 0ADJ: Only with OVC off
Equipment Calibration point Manufacturer Standard model
Voltmeter
0 : 0 V
HIOKI Equivalent to 3237
1 : 1 V
Output cable
Wiring resistance: 500 m or less
(circuit resistance)
Range Calibration point OVC
Measure
ment
Current
100 M
range high-
precision
mode
0ADJ
Resistance
measurement
(Low Power: OFF)
10 m 0 , 10 m ON, OFF - -
With or
without
*1
100 m 0 , 100 m ON, OFF High, Low -
With or
without
*1
1 0 , 1 ON, OFF High, Low -
With or
without
*1
10 0 , 10 ON, OFF High, Low -
With or
without
*1
100 0 , 100 ON, OFF High, Low -
With or
without
*1
1000 0 , 1 k ON, OFF - -
With or
without
*1
10 k 0 , 10 k OFF - - -
100 k 0 , 100 k OFF - - -
1000 k 0 , 1 M OFF - - -
10 M 0 , 10 M OFF - - -
100 M 0 , 100 M OFF - ON, OFF -
1000 M 0 , 1000 M OFF - OFF -
Resistance
measurement
(Low Power: ON)
1000 m 0 , 1 ON - - -
10 0 , 10 ON - - -
100 0 , 100 ON - - -
1000 0 , 1 k ON - - -
Temperature
(thermistor)
25°C: 2186.0
input
Temperature
(analog input)
10°C: 0.1 V input
100°C: 1 V input
D/A output 1
0 : 0 V input
1 : 1 V input