37 ST 800 Pressure Transmitter User’s Manual Revision 15
Note: Sets loop output to low
saturation when the
differential pressure is below -
10% Flow.
Breakpt (%Flow)
(Available only
when Transfer
Function is
Square Root)
##. #
Enter the low flow cutoff
point when Single Breakpt or
Zero Flow is selected. Range:
0.0 to 25.0 %Flow.
Press to enter
menu selection. Press
and to select
number. Press to
enter and shift to
next
digit.
Enter LRV
<Return>
Enter LRV
###. ##
The limit for the Lower Range
Value is 2X the Lower Range
Limit (LRL) of the Meter Body
Enter URV
<Return>
Enter URV
###. ##
The limit for the Upper Range Value
is 2X the Upper Range
Limit (URL) of the Meter Body
Set LRV
<Return>
Set LRV
ATTENTION: Executing this service will set the
Lower Range Value (LRV) equal to the input pressure.
Press to
enter menu
selection
to execute
Set URV
<Return>
Set URV
ATTENTION: Executing this service will set the
Upper Range Value (URV) equal to the input pressure.
Press to
enter menu
selection
to execute
HART Setup
<Return>
Device ID
Unique for each device
Read Only
Universal Rev
HART Revision
Read Only
Field Device Rev
For DD/DTM compatibility
Read Only
Final Assy Num
Asset tracking number
Loop mA
Disabled for Multidrop
Poll Address
0 (default) to 63
PV Units
Units of transmitted PV
SV Units
Units of transmitted SV
HART Date
<Return>
Year
# # # #
Enter the current year.
Month
January -
December
Select the current month.
Day
# #
Enter the day of the month.
Write Date
Press ENTER to write the HART Date to the
transmitter.