Pacer Operating Controls
8-2
8Pacer Operating Controls
The pacer operating controls on the HP CodeMaster 100 are detailed in “Pacer
Operating Controls” on page 1-11.
While pacing, avoid touching the gelled area of the multifunction defib
electrodes or the patient to prevent electrical shock.
Use only Hewlett-Packard recommended electrodes with the external
pacer. The HP CodeMaster 100 delivers pacer pulses through a low-
impedance multifunction defib electrode. The HP CodeMaster 100 does not
pace effectively with high-impedance, pace-only electrodes.
Do not use the same defib electrodes for more than eight hours of
continuous pacing.
8Using the Pacer
To use the pacer, perform the following steps:
1 Apply the multifunction defib electrodes as instructed on the package.
2 Attach the ECG monitoring electrodes as instructed in “Preparing the Leads for
Monitoring” on page 5-3.
3 Attach the patient cable to the ECG Input connector of the HP CodeMaster 100.
4 Attach the patient cable leads to the ECG monitoring electrodes.
5 Attach the electrode adapter cable (M2472A) to the defibrillator output connec-
tor. Pull the latch connector toward the front of the defibrillator to lock the con-
nector in place.