Configuring the CodeMaster 100
2-8
2Configuring the CodeMaster 100
Perform the following steps to configure the CodeMaster 100.
1 Depress Sync and HR Alarm while turning the Energy Select control from Off
to Monitor On. A Setup/Diagnostic menu will appear with the following
choices.
CALIBRATE DEFIB
SETUP MENU 1
SETUP MENU 2
RESTORE FACTORY SETTINGS
PRINT LOG
TEST DEFIB
TEST ECG
TEST CRT
TEST RECORDER
TEST CONTROLS
TEST INDICATORS
TEST BATTERY
TEST PACER
TEST 12-LEAD
2 Select SETUP MENU 1 by pressing the and arrows on the ECG Size
key until the highlight appears on
SETUP MENU 1
.
3 Press Lead Select. SETUP MENU 1 will appear with the current setup values
displayed.
4 Press the and arrows on the ECG Size key until the highlight appears
on the value you wish to change.
5 Press Lead Select.
6 Press ECG Size to scroll through the available choices for this parameter.
7 Press Lead Select to set your selection.
8 Press ECG Size until the highlight appears on the next value you wish to
change.