Delivered Energy and Shock Button Functional Test
To check the HP CodeMaster 100 with the paddles, perform the following
steps:
The HP CodeMaster 100 is designed to protect against delivering a shock
when only one switch on the paddle set is pushed.
1 Turn the Energy Select control to the 100 joules position.
2 Verify that the adult paddle electrodes are installed.
3 Push the paddles completely into their holders (Apex paddle in right
pocket, Sternum in left) and press either Charge. Wait for the Charge
Done indicators.
Keep hands clear of the paddle electrode edges. Use your thumbs to
depress the Shock ( ) buttons on the paddle handles.
4 With the paddles in their holders, grasp the paddle handles and press the
Apex paddle Shock ( ) button. Verify that the defibrillator does not
discharge.
5 Release the Apex paddle Shock button, then press the Sternum paddle
Shock button. Verify that the defibrillator does not discharge.
6 Press Sync to place the defibrillator in sync mode.
7 Press and hold both Shock buttons. Verify that the defibrillator does not
discharge.
8 Press Sync again to remove the defibrillator from sync mode.
9 With the paddles in their holders, press and briefly hold both Shock but-
tons at once. Verify that the defibrillator discharges.
10 The recorder will print a test report. Review the test report for proper
operation.