2-1
2Setup and Configuration
This chapter provides steps for setting up and configuring the HP CodeMaster 100.
Your HP CodeMaster 100 ships protected within a carry bag. For instructions on s
how to attach the side-pouches and shoulder strap, refer to M3710A Carry Bag
Instructions for Use (M2475-94000).
2Setting Up the CodeMaster 100
The HP CodeMaster 100 is ready for operation after you perform the following
tasks:
Insert the Battery.
Load Recorder Paper.
Seat and lock the paddles or defib electrode cable.
Connect the patient cable.
Select configuration settings for both the CodeMaster 100 and the 12-Lead option
including such parameters as;
-date and time
-12-Lead keyboard to uppercase or lowercase mode
-12-Lead telephone directory for transmission
-12-Lead acquisition information
-12-Lead report content information
-12-Lead location and ID codes