Using Shock Advisory Mode
4-4
Verify Patient Condition
Confirm that the patient is:
unresponsive
breathless
•pulseless
A. Select Advisory On
Turn the Energy Select/Power Control to Advisory On to turn on the defibrillator
and operate it in advisory mode. The monitor displays:
FOR ANALYSIS, 200J
PRESS ANALYZE SELECTED
where 200J is the first energy setting. You can configure the energy settings for the
first, second, and third shocks by using the Setup Menu (refer to Chapter 2, “Setup
and Configuration”.
You must turn the Energy Select/Power Control to Advisory On before analyzing
the ECG. Pressing Analyze has no effect unless the Energy Select/Power On
control is set to Advisory On.
In advisory mode, you can use the HP CodeMaster 100 to:
analyze the patient’s cardiac rhythms by pressing Analyze to determine whether a
defibrillation shock is needed
record events with the advisory event summary