6S2
0RQLWRULQJ
SpO
2
Alarms
9-10
9SpO
2
Alarms
Activating SpO
2
Alarms
There are three preset high/low SpO
2
alarm limits: 100/90, 100/85 and 100/80.
Press SpO
2
alarm repeatedly to cycle through the alarm options and the no-
alarm option. Stop when you see the alarm you would like to choose and
after three seconds that alarm will take effect. A symbol replaces the
limits to show that the alarm is active. To review the limit set, press SpO
2
alarm.
If the SpO
2
level falls below the low alarm limit, an alarm sounds and the
violated limit is highlighted.
SpO
2
alarms are automatically turned off when you press Charge.
Deactivating SpO
2
Alarms
Press SpO
2
alarm. The symbol to the right of the SpO
2
display
indicates that the alarms are off.
Recorder Output
After an alarm event, the recorder prints a strip. The bottom of the strip
shows the alarm violation, and the top of the strip shows the SpO
2
reading.
You may print an event summary record, which contains SpO
2
information,
as described in “Printing the Event Summary Record” on page 5-12.