CodeMaster 100
13-8
7 Turn on the defibrillator and press Record to verify the paper is loaded
properly.
Cleaning the Recorder Printhead
If the ECG strip has light or varying density printing, clean the printhead to
remove any possible buildup of paper residue.
To clean the printhead, perform the following steps:
1 Slide the recorder door to the right of the defibrillator. The paper platen
will tilt up.
2 Remove the paper roll.
3 Clean the printhead surface (above the brush) with rubbing alcohol and a
cotton swab.
4 Re-install the paper roll.
5 While holding the recorder door open (to the right), press the platen
down over the paper.
6 Allow the door to close over the platen roller.
7 Press Record to test the printer functionality.