0DLQWHQDQFH
CodeMaster 100
13-4
Notify Service Personnel if the ECG strip does not print TEST 100J
PASSED or if any of the shock button tests fail.
If you see the CHECK SETUP message on the screen, call service to replace
the backup battery.
Electrode Adapter Cable Test
To check the electrode adapter cable, perform the following steps:
1 Connect the HP M1781A test load to the electrode adapter cable.
2 Turn Energy Select to 100 joules.
3 Press Charge. Wait for the Charge Done indicators.
4 Press the left Shock ( ) button on the pads adapter cable connector
housing. Verify that the defibrillator does not discharge.
5 Release the left Shock button and press the right Shock button on the
pads adapter cable connector housing. Verify that the defibrillator does
not discharge.
6 Press Sync to place the defibrillator in sync mode.
7 Press and hold both Shock buttons. Verify that the defibrillator does not
discharge.
8 Press Sync again to remove the defibrillator from sync mode.
9 While holding down Charge, press both Shock buttons on the pads
adapter cable connector. Verify the defibrillator discharges.
The recorder will print a test report.
NOTE
CAUTION