0DLQWHQDQFH
CodeMaster 100
13-9
Cleaning Exterior Surfaces
The HP CodeMaster 100 and accessories (except the ECG cable and carry
bag) are chemically resistant to common hospital cleaning solutions and
non-caustic detergents. The following list includes some approved cleaning
solutions:
90% Isopropyl alcohol (except adapters and patient cable)
Soap and water
Chlorine bleach (30 ml/l water)
The following list includes practices for keeping the HP CodeMaster 100
exterior surfaces clean:
Keep the outside of the instrument clean and free of dust and dirt. Clean the
paddles thoroughly to prevent build-up of dried electrolyte medium.
Do not allow any fluids to penetrate the instrument case. Avoid pouring
fluid on the unit while cleaning.
Do not use abrasive cleaners, or strong solvents such as acetone or ace-
tone-based compounds.
Clean the display screen carefully. It is especially sensitive to rough
handling and subject to scratching.
Do not steam sterilize the monitoring leads, submerge them for prolonged
periods, or heat them above 50°C. If metallic surfaces become oxidized,
clean them with a very light abrasive (toothpaste). Do not use highly
abrasive cleaners such as steel wool or silver polish.
Do not steam or gas sterilize the external paddle set.