6SHFLILFDWLRQV
Defibrillator/Monitor
14-3
Monitor
Inputs: ECG may be viewed through paddles or patient cable. Lead I, II, III, or
PADDLES selectable. Additional leads (aVR, aVF, aVL, V Leads) and PADS are
available. Monitor and recorder indicate selected ECG source.
Lead Fault: LEADS OFF message and dashed baseline appear on monitor if a lead
becomes disconnected.
Common Mode Rejection: Greater than 90 dB measured as per AAMI standards
for cardiac monitors (EC13).
Display Size and Type: 5 inch (12.7 cm) diagonal CRT for 4 seconds of ECG data
on screen; non-fade, fixed trace. Scrolling trace is selectable.
Sweep Speed:
-Fixed trace: 25 mm/sec nominal
-Scrolling trace: 28.5 mm/sec nominal
Frequency Response: 0.5 to 40 Hz.
Heart Rate Display: Digital readout on monitor from 15 to 300 BPM.
Heart Rate Alarms: Three configurable pairs of high and low heart rate alarm
limits from 20 to 280 BPM.
ECG Output: 1V/mV.
Patient Cable Length: 10 ft (3 m).
External Pacer
Current Pulse Amplitude: 10 mA to 200 mA.
Pulse Width: 20 msec.
Rate: 40 ppm to 180 ppm.
Modes: Demand or fixed rate.
Refractory Period: 340 msec (40 to 80 ppm); 240 msec (90 to 180 ppm).