Configuring the CodeMaster 100
2-10
To set the time on the CodeMaster 100:
1 Press Sync and HR Alarm at the same time while turning the Energy Select
control from Off to Monitor On.
2 Press ECG Size once to highlight SETUP MENU 1.
3 Press Lead Select. SETUP MENU 1 appears.
4 Press ECG Size seven times to highlight Time.
5 Press Lead Select.
6 Press ECG Size until the hours setting contains the correct time.
7 Press Lead Select.
8 Press ECG Size to select minutes.
9 Press Lead Select.
10 Press ECG Size until the minutes setting contains the correct time.
11 Press Lead Select.
12 Depress both sides of the ECG Size key simultaneously to return to the
Setup/Diagnostic Menu.
13 Turn the Energy Select control to Off.