Defibrillating a Patient
3-8
If the PADS OFF monitor message displays, check all patient connections.
5 Select pads as the ECG source by pressing Lead Select until PADS appears on
the display under the heart rate.
6 Set the Energy Select control to desired energy.
7 Press Charge.
8 Call out "Clear!" to alert personnel to stand away from the patient.
9 Wait for the charge done indicators.
10 Press both Shock buttons simultaneously to defibrillate. The Shock () but-
tons for the multifunction defib electrodes are on the cable connector housing.
(Figure 3-2).