Setup mode
Chapter 4 OPERATION MANUAL
V1.3.0en/06.07.15//14.06 Liability for errors and misprints excluded. CC®
4.1.3.3 Cooling/heating baths: Switching K1x and K2x on (with hanger thermostat)
The cold bath in combination with a hanger thermostat must only be used in switch position "II" -
controller operation (on the cooling bath). The fuse may be tripped by excessive current consump-
tion if ignored.
PROCEDURE
Switch on the cooling bath using the >mains switch< [37] (Position “II” - controller operation).
Switch on the hanger thermostat using the >mains switch< [37] (Position “I”).
First, a system test is conducted to test the full functionality of the temperature control unit. All
sensors, the all pole breaking power disconnect relay for the power supply and the power elec-
tronics of the main heater and the main heater itself are checked. A message appears on the
>touchscreen< [88] in case of an error or if a warning is issued. If in doubt contact Customer
Support (the phone number can be found on page 82 in section »Phone number and company
address«).
The following data entry is required only for:
a) initial startup
b) reseting the temperature control unit to the factory settings
(see page 43, section »Restore factory settings«).
After switching on the temperature control unit, tap on the required system language.
Confirm your choice by tapping on “OK”.
Tap on the thermal fluid used.
Confirm your choice by tapping on “OK”.
Read the message and confirm by tapping on “OK”.
Enter the volume using the number keypad that appears.
Confirm your entry by tapping on “OK”.
Tap on the cooling bath used. This selection list is displayed only when the Pilot ONE was con-
nected to a cooling bath.
Confirm your choice by tapping on “OK”.
If your temperature control unit is not integrated into a network, confirm the preset IP address
(0.0.0.0) by tapping on “OK”. This skips the network settings.
Enter the required IP address using the number keypad that appears.
Confirm your entry by tapping on “OK”.
Enter the required Subnet Mask using the number keypad that appears.
Confirm your entry by tapping on “OK”.
Tap on the required remote control mode.
Confirm your choice by tapping on “OK”.
4.1.3.4 Cooling/heating baths: Switching K1x and K2x off (with hanger thermostat)
PROCEDURE
Switch off the hanger thermostat using the >mains switch< [37] (Position “0”).
There is no need to switch off the cooling bath. Leave the >mains switch< [37] on Position “II” -
controller operation. If the temperature control unit is to be switched of for an extended period,
set the >mains switch< [37] on the cooling bath to position "0” - Off.