Setup mode
Chapter 4 OPERATION MANUAL
V1.3.0en/06.07.15//14.06 Liability for errors and misprints excluded. CC®
61
4.2.9.1.3 Sub-category: “Process
Enter the new values for “KP”, “Tn” and “Tv” here one after the other.
PROCEDURE
Go to the “Categories Menu”.
Tap on the category “Temperature Control”.
Tap on the category “TAC/Manual”.
Tap on the sub-category “Configuration manual”.
Tap on the sub-category “Change parameters”.
Tap on the sub-category “Process”.
Enter the new “KP” value using the number keypad that appears.
Confirm your entry by tapping on “OK”.
Enter the new “Tn” value using the number keypad that appears.
Confirm your entry by tapping on “OK”.
Enter the new “Tv” value using the number keypad that appears.
Confirm your entry by tapping on “OK”.
4.2.9.2 Sub-category: “Display parameters”
In this function, the set manual parameters are displayed.
PROCEDURE
Go to the “Categories Menu”.
Tap on the category “Temperature Control”.
Tap on the category “TAC/Manual”.
Tap on the sub-category “Configuration manual”.
Tap on the sub-category “Display parameters”.
Tap on “OK” after you have read/checked the parameters.
4.2.9.3 Sub-category: “Control structure”
With this function, you have two different control structures available.
„Huber PID controller“: Default setting
“Classic PID controller”: This setting is used by the service engineers of Peter Huber Kältemaschi-
nenbau GmbH exclusively for service purposes.
PROCEDURE
Go to the “Categories Menu”.
Tap on the category “Temperature Control”.
Tap on the category “TAC/Manual”.
Tap on the sub-category “Configuration manual”.
Tap on the sub-category “control structure”.
Choose between the dialogue entries “Huber PID controller” and “Classic PID controller”.
Confirm your choice by tapping on “OK”.
4.2.10 Sub-category: “Reset parameters”
With this function, you can reset the control parameters to the factory setting.
PROCEDURE
Go to the “Categories Menu”.
Tap on the category “Temperature Control”.
Tap on the category “TAC/Manual”.
Tap on the sub-category “Reset parameters”.
Read the message and confirm it by tapping on “Yes”. The control parameters are reset/deleted.
The temperature control unit can only be operated again after a complete restart.
To do so, switch the temperature control unit off and back on again. The parameters have been reset.