Normal operation
Chapter 5 OPERATION MANUAL
V1.3.0en/06.07.15//14.06 Liability for errors and misprints excluded. CC®
67
5 Normal operation
5.1 Automatic operation
Hot or cold thermal fluid and surfaces
BURNS TO LIMBS
Avoid direct contact with the thermal fluids or the surfaces.
Wear your personnel protective equipment (e.g. temperature-resistant safety gloves, safety
goggles, safety footwear).
5.1.1 Temperature control
5.1.1.1 Starting the temperature control process
The temperature control process can be started after filling and complete venting.
PROCEDURE
Go to the “Home” screen.
Tap on the “Start” touchbutton.
Confirm the start of temperature control by tapping on “OK”.
The correct selection will be displayed graphically and temperature control will start immediate-
ly. If tapping on “OK” is not correct, this is displayed graphically for 2 seconds. After this, the dis-
play will return to the “Home” screen again. Try to start the temperature control unit again.
5.1.1.2 Ending the temperature control process
When the temperature control unit is switched off, the thermal fluid temperature is high-
er/lower than room temperature
DAMAGE TO THE TEMPERATURE CONTROL UNIT AND THE GLASS APPARATUS/APPLICATION
Bring the thermal fluid up to room temperature using the temperature control unit.
Do not close the shut-off valves in the thermal fluid circuit.
The temperature control can be ended at any time, the pump continues to run for approx. 30 sec-
onds. The compressor is switched off after the stepper motor valve has reached a defined position
to control the cold output.
PROCEDURE
Go to the “Home” screen.
Tap on the “Stop” touchbutton.
Confirm the stop of temperature control by tapping on “OK”.
The correct selection will be displayed graphically and temperature control will stop immediately
and the pump with overrun for about 30 seconds. Wait until the pump stops. If tapping on “OK”
is not correct, this is displayed graphically for 2 seconds. After this, the display will return to the
“Home” screen again. Try to stop the temperature control unit again.
The compressor is not switched off until the stepper motor valve has reached a defined position.
The status line [Field 10] displays the relevant information.