Service/maintenance
Chapter 7 OPERATION MANUAL
V1.3.0en/06.07.15//14.06 Liability for errors and misprints excluded. CC®
79
7.6 Cleaning the surfaces
Exposed plug contacts
DAMAGE CAUSED BY FLUID INGRESS
Protect unused plug contacts with the protective caps supplied.
Clean surfaces only with a damp cloth.
A standard stainless steel cleaning agent is suitable for cleaning the stainless steel surfaces. Carefully
clean paint surfaces (damp) using the suds of a mild detergent.
7.7 Inspect the mechanical seal
No visual inspection of the mechanical seals
MATERIAL DAMAGE IN THE TEMPERATURE CONTROL UNIT CAUSED BY LEAKING MECHANICAL
SEALS
Check the mechanical seals once a month.
If case of leakage, stop the temperature control unit and contact Customer Support. The tele-
phone number can be found on page 82, section »Phone number and company address«.
Expect drop formation at the mechanical seal when operating with thermal fluids that evaporate
only very slowly, as mechanical seals are never absolutely tight. These drops must be removed if
necessary (see page 74, section »Function check and visual inspection«). The tightness of the me-
chanical seal must be visually checked. In case of a leakage, more thermal fluid exits at the bottom
of the temperature control unit.
7.8 Plug contacts
Exposed plug contacts
DAMAGE CAUSED BY FLUID INGRESS
Protect unused plug contacts with the protective caps supplied.
Clean surfaces only with a damp cloth.
Protective caps are supplied for all plug contacts. Make sure that any plug contacts not required are
protective with the caps.
7.9 Decontamination/repairs
Returning a not decontaminated temperature control unit for repair
PHYSICAL INJURY AND PROPERTY DAMAGE CAUSED BY HAZARDOUS MATERIALS IN OR ON THE
TEMPERATURE CONTROL UNIT
Carry out appropriate decontamination.
The decontamination process depends on the type and quantity of the materials used.
Consult the relevant safety data sheet.
You will find a prepared return receipt at www.huber-online.com.
You as the responsible body are responsible for carrying out decontamination BEFORE third-party
personnel come into contact with the temperature control unit. Decontamination must be carried
out BEFORE the temperature control unit is returned for repair or inspection (clearly stating in writ-
ing on the temperature control unit that decontamination has been carried out).
To simply the process, we have prepared a form for you. This is available for download at
www.huber-online.com.